Cloud Condensation Nuclei Counter with Single (CCN-100) or Dual (CCN-200) Growth Columns

 

Cloud Condensation Nuclei Counter with Single (CCN-100) or Dual (CCN-200) Growth Columns

data icon  View Sample Realtime Data Display

download icon Download Sales Brochures:
              Cloud Condensation Nuclei Counter (CCN-100)

              Cloud Condensation Nuclei Counter, Dual Column (CCN-200)

              CCN Inlet Pressure Controller

 

 

DMT SPECIFICATIONS:

 

  • Measures cloud condensation nuclei (CCN) using a supersaturated column and optical particle counter

  • Continuous flow, single supersaturation

  • Supersaturation as low as 0.07%

  • Cylindrical geometry reduces size and buoyancy effects

  • Vertical thermal gradient

  • Continuous wetting

  • Multiple bin optical particle counting (20 size bins)

  • Onboard computer control and data logging

  • Set up for airborne or ground-based applications

  • 28 VDC power, 15 A

  • Rack mountable, 28 kg weight (single column) or 38 kg weight (dual column)

  • System includes CCN instrument, test cables, shipping case, software, one day of training at DMT's facility, and one-year warranty

  • Technique has been developed and patented by The University of California, San Diego, and licensed for production by DMT

CCN Software:



The CCN comes with customized software, which allows users to configure the instrument settings quickly and easily. For instance, the program can be used to:


  • Change supersaturation settings

  • Change temperature settings

  • Change the sample air flow

  • Manipulate instrument pumps (e.g., turn air pumps on high to prevent condensation)

  • Quickly detect any operational problems

  • Update instrument calibration parameters

  • Adjust the instrument to prepare it for shipping or re-humidify it after shipping


  • CCN software also performs data collection. Information gathered during sampling sessions is written to output files that can be viewed real-time and played back later for detailed analysis.


    In addition to this instrument-specific software, the CCN interfaces with DMT's Particle Analysis and Display System (PADS). PADS allows the user to analyze data collected from the CCN and other DMT instruments simultaneously.

     
    Registered users, log in to see specialized content. (For current customers only; contact DMT to register.)