The complete CCN Cleaning and Calibration includes the following:
|
Incoming unit evaluation and calibration check (if unit is functional)
OPC cleaning and calibration
Bisque visual inspection and disinfection
Temperature readings and set-point calibrations
Nafion membrane replacement
Sheath flow filter replacement
Sample manifold and sample capillary cleaning
Flow and pressure calibration
DMA comparison and temperature versus SS% calibration
100% unit functionality test (data ports, Ethernet, etc.)
Computer battery replacement
|
|
The customer receives copies of the calibration data and report.
DMT technicians check and verify the CCN in for service in a similar manner in which they verify a new CCN.
|